Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8021B: CAL ADJ /5500 INSTRUMENT: Fluke 8022B: CAL ADJ /5500 DATE: 16-Jan-95 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 73 CONFIGURATION: Fluke 5500A ============================================================================= # # Source: # Fluke 8021B instruction manual (PN 616037 July 1981). # Fluke 8022B instruction manual (PN 616045 July 1981). # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 55_v8021.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5500A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- J F W 1.002 HEAD EQUIPMENT SETUP 1.003 DISP [32] WARNING 1.003 DISP HIGH VOLTAGE is used or exposed during the performance 1.003 DISP of this calibration. DEATH ON CONTACT may result if 1.003 DISP personnel fail to observe safety precautions. 1.004 DISP Ambient temperature: 21C - 25C. 1.004 DISP Relative humidity: less than 80%. 1.004 DISP Allow the UUT to stabilize for 30 minutes 1.004 DISP with the power switch off. 1.005 HEAD CALIBRATION ACCESS PROCEDURE 1.006 DISP Disconnect the test leads and the battery eliminator 1.006 DISP if attached. Open the battery compartment, and 1.006 DISP disconnect the battery. Remove the three phillips-head 1.006 DISP screws from the bottom of the case. Turn the instrument 1.006 DISP face-up, and grasp the top cover at both sides of the 1.006 DISP input connectors. Then, pull the top cover from the UUT. 1.007 DISP Re-connect the battery and turn the UUT on. 1.008 HEAD DC VOLTAGE CALIBRATION 1.009 PIC 55_V8021 1.010 RNG 200mV 1.011 5500 190mV S 2W 1.012 DISP Adjust R5 for a reading of 190.0mV exactly. 1.013 5500 * S 1.014 HEAD AC VOLTAGE CALIBRATION 1.015 RNG 200mV AC 1.016 5500 190mV 100H SI S 2W 1.017 DISP Adjust R9 for a reading of 190.0mV (+/- 1/2 count). 1.018 END